Other input/outputs: Filter Code (2 bits) does not change and Filter Change output is not used, in normal operation (disabled).
Details:
PC interface for Argentine Airglow Spectrometer
1. Address assignment: 238-23F = base address 238 + x, x=0,...,7
2. Port assigment:
Port # bit INPUT OUTPUT
----------------------------------------------------------------------------
0 D0 D0 \ timer
1 D1 D1 select
2 D2 D2 / ("CO")
238 3 D3 LSB D3 PM HiVolt relay
4 D4 photon D4 counter reset
5 D5 counter D5 stepping motor forw/reverse
6 D6 D6 relay 2 (filter change)
7 D7 D7 relay 3 (shutter)
----------------------------------------------------------------------------
0 D0
1 D1
2 D2 stepping motor
239 3 D3 MSB pulse
4 D4 counter (1 bit)
5 D5
6 D6
7 D7
----------------------------------------------------------------------------
0 D0 detector temp. T1 D0 relay 4
1 D1 upper limit sw D1
2 D2 lower limit sw D2
23A 3 D3 P2 filter.. D3
4 D4 P1 ..pos D4
5 D5 available D5
6 D6 count ready D6
7 D7 filter temp. T2 D7
----------------------------------------------------------------------------
0 D0 counter carry bit0
1 D1 counter carry bit1
2 D2 counter overflow
23B 3 D3 available
4 D4 available
5 D5 available
6 D6 available
7 D7 available
--------------------------------------------------------------------------